Probe → Zone Mapping

Each physical thermocouple probe must be mapped to a zone (1-6). Heat-detect protocol (mirrors scripts/identify_tcs.py): pick a zone, Capture Baseline, grip the probe you want for that zone (skin contact for ~15 s), then Detect. The probe with the largest rise (≥ 2°C above baseline) is auto-assigned. Repeat for each zone, then Save Mapping. Saved mappings take effect within ~10 seconds (next sensor loop cycle).

Detect zone:
Probe ID Live Temp Δ since baseline Zone
Scanning 1-Wire bus…

Temperature Probes

2-Point Cal (recommended): record two reference temperatures (e.g. ice bath 0°C and hot water 50°C) for a linear correction. Quick Cal: single-point offset using one reference. Both methods auto-apply — no service restart needed.

Zone Live Reading Calibration Status Actions

Load Cells

Tare: clear the scale, click Tare. Calibrate: place a known weight, type its mass in kg, click Calibrate. The running scale picks up the new tare/factor automatically — no service restart needed.

Calibrate All (centered load)

Empty the table, click Tare All. Then place a known total weight at the center of the table and enter its mass — we split equally across the 4 corner cells.

Step 1:
Step 2: Total weight: kg (each scale gets total / N kg)
Scale Live (kg) Raw Tare Factor Known kg Actions